[Video] Starting Your Compounding Aseptic Isolator

January 9, 2017

Prior to use, the PharmaGard 797 barrier isolator must be cleaned. Press the On button to start the blower and illuminate the lights.

Product testing of the PharmaGard 797 has confirmed that ISO Class 5 conditions are achieved within one minute of activating the blower. NuAire has recommended waiting 5 minutes before accessing the work zone to ensure it is particle free.

Check the magnihelic gauges on the control panel to ensure all pressure readings are within product specifications.

Contact a supervisor or NuAire if you encounter and issue starting up your Pharmagard 797.