[Report] CO2 Incubator Decontamination Validation Report

October 11, 2016

CO2 Incubator Decontamination Validation ReportTo help minimize the risk of CO2 Incubator contamination, advanced models of In-VitroCell Direct Heat CO2 Incubators feature a dry and humidified decontamination cycle to eliminate unwanted growth in the chamber that pose a threat to your cell lines. This report validates the performance of a 145°C dry heat decontamination cycle using CO2 Incubator NU-5800 series model NU-5840 showing a total inactivation of a biological indicator with a log10 population of 6.6. 